HI3820 มีย่านการทดสอบที่กว้างได้แก่ 0 ถึง 100 mg/L ด้วยความละเอียด 1 mg/L และ 0 ถึง 500 mg/L ด้วยความละเอียด 5 mg/L ชุดทดสอบสามารถใช้งานได้ประมาณ 110ตัวอย่าง
Specifications HI3820
SKU | HI3820 |
---|---|
Product Name | Acidity Chemical Test Kit |
Acidity Measurement Type | titration |
Acidity Range | 0 to 100 mg/L CaCO₃; 0 to 500 mg/L CaCO₃ |
Acidity Resolution | 1 mg/L (0 to 100 mg/L range); 5 mg/L (0 to 500 mg/L range) |
Acidity Method | methyl orange / phenolphthalein |
Acidity Number of Tests | 110 |
Weight | 910 g (34.0 oz.) |
Dimensions | 260x120x60 mm (10.2x4.7x2.4") |
Ordering Information | HI3820 test kit comes with 1 (10 mL) dropper bottle dechlorinating reagent, 1 (10mL) dropper bottle bromophenol blue indicator, 1 (10mL) dropper bottle phenolphthalein indicator, 1 (120mL) bottle HI3820-0 acidity titrant, 10 mL calibrated vessel, 50 mL calibrated vessel, and calibrated syringe. |
หมายเหตุ
ความแตกต่างระหว่างการวัดค่า pH และ Tritratable acidity
- ค่า pH เป็นส่วนกลับของลอการึทึม (logarithm) ของความเข้มข้นของไฮโดรเจนไอออน (H+) เป็นกรัมต่อสารละลายหนึ่งลิตร โดยค่า pH แสดงความเป็นกรดหรือเบสของสารละลาย มีค่าตั้งแต่ 0 ถึง 14 โดยค่า pH มีค่า 7 หมายความว่ามีความเป็นกลาง (natural pH) pH มีค่าต่ำกว่า 7 แสดงความเป็นกรด (acidic pH) pH มีค่าสูงกว่า 7 แสดงความเป็นเบส (alkaline pH)
- ส่วน Titratable acidity เป็นการวัดปริมาณกรดอินทรีย์หลักที่มีอยู่ในตัวอย่างเช่น น้ำผลไม้ นม ไวน์ วัดด้วยการไตรเตรท ด้วยด่าง (NaOH) โดยผลการไตเตรทบอกปริมาณกรดอินทรีย์ที่มีอยู่หลักในตัวอย่าง เช่น ในไวน์ มีกรด tartaric เป็นหลัก แสดงเป็น g/L tartaric acid equivalent หรือ ในนม มีหน่วยเป็น g/L lactic acid equivalent
PDF Catalog
Titration is used to find the concentration of an identified chemical in a solution (e.g., water).
In our standard Titration kits, there will be a Titrant solution that is added to the liquid (e.g., water) to be tested to make it change color. The Reagent is then added under monitored conditions.
Acidity Test Kit
Acidity can be caused by mineral or organic acids, or by carbon dioxide as Carbonic Acid. Acidity of aqueous solutions contributes to corrosion and influences the rate of biological processes.
The HI-3820 acidity test kit makes monitoring quick, easy and safe. During testing, when colour change is achieved, under careful conditions – usually using a burette – the reagent is added, drop by drop, until the solution turns to the colour required.
The user will then count how many drops of reagent were used to achieve complete colour change. The number of drops used will determine the chemical concentration and results can be found in the manual included.
However, the compact size gives the user the versatility to use this kit anywhere.
With the use of diluted sodium hydroxide as the titrant and bromophenol blue or phenolphthalein indicators, the contribution of strong or organic acids can be determined.
The measurement of the strong acid contribution to the sample acidity is known as methyl orange acidity. This is carried out by titrating with sodium hydroxide until the solution turns from yellow to green/blue (pH endpoint about 4.5). T
he total acidity caused by both mineral and organic acids is determined by titrating to an endpoint pH of 8.3, using phenolphthalein as an indicator. This is known as phenolphthalein in acidity.